Candela Error Codes
Service Resource / Error Codes by Brand / Candela
GentleMax Pro and GentleMax Pro Plus
45 faults and 4 warnings, grouped for quick reference. These mappings agree in the two reviewed revisions; coverage of newer software is not established.
GentleMax Pro: 8501-00-2200 Revision D, pages 110–120.
GentleMax Pro Plus: 8501-00-2400 Revision D, April 21, 2022, pages 102–111.
Power, shutter and calibration
F3.1 — Shutter position/actuation failure
F4.2 — Supply communication timeout
F5.1 — Supply outside tolerance
F5.2 — Charging timeout
F5.3 — Discharge-test failure
F6.1 — Calibration energy not reached
F6.5 — Calport switches disagree
Water cooling and warm-up
F7.1 / F7.2 — Water too cold / hot
F7.3 — Circulation or pressure problem
F7.4 — Water-temperature sensor fault
F7.5 — Insufficient water
F9.1 / F9.2 — Water / canister warm-up failure
Cryogen and DCD
F1.1 / F1.2 — Handpiece / canister bubble-sensing test failure
F8.2 — Excessive canister pressure
F8.3 — Handpiece spray-valve fault
F8.4 — Canister temperature-sensing fault
Energy, pulse timing and optical delivery
F10.1 — Slider identification failure
F12.1 / F12.2 — Pulse energy below / above range
F12.3 — Energy ceiling exceeded
F12.4 — Unequal subpulse energy
F12.6 — Pulse duration outside tolerance
F14.1 / F14.2 — 755 / 1064 nm simmer fault
F15.1 / F15.2 — Transmission below / above range
Trigger controls
F13.1 / F13.2 — Finger / foot trigger channels disagree
F13.3 / F13.4 — Finger / foot trigger remains activated
Circuit, software and communication faults
F18.1 / F18.2 / F18.3 — Energy / water / DCD calibration fault
F19.1 — Triggering fault
F19.3 / F19.4 — Laser / head power fault
F19.5 — Discharge-test failure
F21 — Incomplete update
F23 — One-wire network fault
F24 — Interface communication fault
F26.1 / F26.2 — Internal software fault
Warning codes
10.2 — Delivery assembly disconnected in Ready
10.3 — Slider removed in Ready
10.4 — Fiber connection absent in Ready
17 — Handpiece cryogen bubbles
Using this reference
Match the exact model, software version, full screen message and applicable manual. A code identifies a detected condition and does not by itself diagnose a failed component. Perform only checks permitted by the exact operator manual and your training. Persistent faults or unresolved concerns require qualified assessment before clinical use resumes.
Request equipment service
Send the model, serial number, full code, handpiece, software version, when the fault occurs and checks already completed. Exclude patient identifiers from photographs.
Contact VMS for equipment service | 833-286-7669
Reference documents
GentleMax Pro operator manual — 8501-00-2200 Rev D
GentleMax Pro Plus operator manual — 8501-00-2400 Rev D
Publicly hosted copies of Candela-authored manuals. Reviewed November 6, 2026. VMS is an independent service provider, not affiliated with or endorsed by Candela. Manufacturer names identify equipment only.