Xeo / Xeo SA / Xeo Pearl

Selected documented codes from Cutera service manual D0117 Rev. B, Section 5, pages 5-18–5-20. These older-platform mappings do not establish coverage for Xeo+, Excel V, Excel HR, Enlighten or newer software.

Xeo — cooling and controls

204 — Safety interlock open

221 — Coolant temperature high

224 — Tip temperature cutoff exceeded

226 — Handpiece parked during exposure

229 — Tip excessively hot

250 — Footswitch fault

251 / 252 — Shutter opening / closing failure

254 — Ready selected while footswitch depressed

400 — Reservoir vacuum fault

401 — Water pump reported off

402 — Water blockage detected

Xeo — energy and flashlamp

230 / 231 — Energy below / above expected level

235 — Flash handpiece efficiency fault

243 — Lamp simmer voltage/current outside limits

246 — Flashlamp ignition failure

248 — Flashlamp simmer absent

249 — Nd:YAG flashlamp ignition failure

361 — Nd:YAG efficiency insufficient

362 — Nd:YAG output regulation fault

Xeo — communication, software and calibration

270 / 272 — Power supply communication timeout

273 — Flash handpiece communication retry fault

363 — System data integrity error

365 — Handpiece startup self-test failure

366 — Nd:YAG calibration incomplete

367 — Power supply startup self-test failure

368 — Touchscreen calibration fault

369 — Test-fire timeout

380 — Flash handpiece communication timeout

386 — Handpiece authorization fault

387 — Flash handpiece calibration incomplete

390 — Software runtime error

392 — Software version mismatch

Excel V

The operator manual D1025 Rev. H, October 2017, page 50, identifies “Connect Footswitch” and “Water Low” messages. Its error guide does not provide a detailed numeric dictionary. Numeric codes 273 and 380 remain unverified for Excel V.

Excel HR

A manufacturer definition has not yet been verified for codes 766 or 767. Record the complete screen message and match the exact model, software version and applicable manual before interpreting a code.

Enlighten

Manufacturer definitions for codes 1902 and 1903 remain unverified. Do not apply Xeo code meanings to this platform.

Entries awaiting verification

Xeo 335 is not confirmed in the reviewed table. The reviewed Xeo document contains conflicting descriptions for 225 and 323; these require the exact screen message and revision before interpretation.

Using this reference

A code identifies a detected condition and does not by itself diagnose a failed component. Perform only checks permitted by the exact operator manual and your training. Persistent faults or unresolved concerns require qualified assessment before clinical use resumes.

Request equipment service

Send the model, serial number, complete screen message, handpiece, software version, when the fault occurs and checks already completed. Exclude patient identifiers from photographs.

Contact VMS for equipment service | 833-286-7669

Reference documents

Xeo service manual — D0117 Rev. B (publicly hosted copy)

Excel V operator manual — D1025 Rev. H, error guide

Reviewed November 6, 2026. VMS is an independent service provider, not affiliated with or endorsed by Cutera. Manufacturer names identify equipment only.

VMS logo

Pages

Legal

Subscribe

Call Us

Stay up to date on everything lasers, news, & exclusive VMS discounts.

Email

  • Instagram
  • Facebook
  • LinkedIn
  • YouTube

Disclaimer: Ventura Medtech Solutions dba Aesmed Solutions is not affiliated, associated, authorized, endorsed by, or in any way officially connected with any equipment manufacturer or their affiliates. Any logos and trademarks of third parties that may be found on our website are registered trademarks of their respective owners. The use of any trademark is for identification and reference purposes only and does not imply any association with the trademark holder.